SWG 100 BIOcompact
simple and economical solution for all small scale biogas applications
sequential measurement o…
High quality flue gas conditioning guarantees highly accurate measuring results. Condensation of the flue gas sample must be avoided in the pathway from the point of gas extraction to the flue gas analyzer. None of the target substances must be lost due to condensation. The gas conditioning system protects the analyzer against humidity and dust particles. Thus, long term measuring over a period of several hours becomes possible.
portable gas conditioning system
Measuring range
| Flow | 50 … 150l/h or 50 … 250l/h (w/ double heat exchanger) |
| Ambient air temperature | +5°C … +50°C |
Dimensions and other data
| Cooling capacity | 100 kj/h |
| Inlet dew point | 40°C up to 65°C max. (depending on flow rate) |
| Outlet dew point | + 5°C, Alarm at deviation of ± 3°C |
| Cooler type | Peltier controlled heat exchanger |
| Power consumption | max. 120W |
| Warm up time | appr. 5 minutes |
| Dimensions | 410x360x210 mm (WxHxD) |
| Weight | 10 kg |